ul. Mickiewicza 59, 47-253 Cisek
+ 48 77 550 72 00

Przewoźny aparat RTG Mobile 320

Specifikations

Mobile 320
X-ray group

Monoblock

Max. power (100 kV – 320 mA): 32 kW
Max voltage at the tube: 125 kV
Max current at the tube: 450 mA
Min. inherent filtration @75kV: 1.1 mmAl
Thermal capacity available: 500 kJ
Continuous thermal dissipation: 55 W
Weight: 19.4 kg

X-ray tube

Rotation speed: 3000 min-¹
Nominal High voltage: 130 kVp
Nominal focus dim. (IEC 60336): 0,8 mm small focus, 1,3 mm large focus
Nominal anodic power (IEC 60613): 16 kW small focus, 32 kW large focus
Thermal capacity of the anode: 80 kJ (107 kHU)
Max continuous anode dissipation: 300 W
Min. inherent filtration (IEC 522): 0,7 mmAl eq.

Collimator

Collimation: Manual with internal light source, multilayer, squared field.
Coverage of the field at 100cm FFD (SID): min 0 x 0 cm, max 43 x 43 cm
Lighting source: Clusters of high-brightness LED power.
Focal distance measurement: Retractable tape measure (max extension 3 m)
Rotation: ± 120°
Weight: 5.5 kg

Mains connection

115 Vac / 230 Vac ±10%, standard monophase with ground conductor. Automatic setting of the equipment according to mains voltage 50/60 Hz ± 5 Hz

Absorbed current

10 A

Line compensation

Automatic

Line resistance

<1 Ω @ 115/230 Vac

Standard socket

16 A @ 230 Vac

Power supply cable

8 m

W x L x H (in transport position)

618 mm x 1366 mm x 1444 mm

Transport handle height

949 mm

Focus-ground distance (Z-axis)

417 mm ÷ 2092 mm

Max. height

2333 mm

Rotation of the monobloc

around the arm axis (α swivel): ± 180°
around its axis (γ swivel): 151° (102° ahead, 49° back)

Movement

manual

Weight

170 kg

X-ray Specifications

Environmental Conditions

Nominal power (IEC 60601-1)

32 kW @100 kV, 320 mA, 100 ms

kV variation range

40 ÷ 125 kV at steps of 1 kV

mA variation range

50 ÷ 400 mA

mAs variation range

0.1 ÷ 220 mAs

Exposure times range

0.001 ÷ 2.2 s according to mAs

Operating
Temperature: 10 °C to 40 °C (50 °F to 104 °F)
Humidity: 30 % to 75 % (non-condensing)
Atmospheric pressure: 700 hPa to 1060 hPa

Transport / Storage
Temperature: -25 °C to 70 °C (-4 °F to 122 °F)
Humidity: 10 % to 90 % (non-condensing)
Atmospheric pressure: 500 hPa to 1060 hPa